The new X-Mind AC generator is fitted with a Toshiba DG-073B monophased auto rectifier tube. It generates a high 70kV voltage with an ideal intensity of 8mA. Concentrated, high density radiation in a very small diameter focal spot (0.7mm) for greater accuracy. Supplied with a long cone as standard, the generator is perfectly suited to the parallel technique (a short cone and a rectangular cone are available as options). The rays emitted have a large wavelength and provide pictures characterized by a widespread grey scale for better distinction of anatomical detail. The new timer is programmed for use with “D”, “E” or “F” type films. It can be installed remotely and can control one or two units whatever the technology (AC + AC or AC + DC), making the X-Mind concept unique and providing the most flexibility of use in intra-oral radiography.
Classification: Electromedical equipment, Class 1 type B
Supply voltage: 220/230/240V~ - monophase 50/60Hz
Power absorption at 230V: 0.8kVA
Resistance of the line: 0.5 ohms
X-ray tube: New Toshiba DG 073B tube
High voltage: 70kV
Anode current: 8mA
Focal spot: 0.7mm
Total filtration: Equivalent to 2mm Al at 70kV
Tube inherent filtration: Equivalent to 0.8mm Al at 70kV
Leakage radiation: Less than 0.25mGy/hour maximum at a distance of 1m from focal spot
Long cone: Focus to cylinder tip distance = 31cm (12”)
Timer: Exposure time can be set from 0.08 to 3.2 seconds
X-ray emission control: “Dead man” switch with 3m spiral cable
Total weight: 28kg (weight of the head: 9kg)
Manufactured in compliance with currently applicable regulations and standards (EC Directive 93/42/EEC)
X-Mind® DC
The new X-Mind DC generator is fitted with a Toshiba DG-073B-DC double anode tungsten tube (intensity of 4mA and 8mA). It generates a constant voltage of 60kV or 70kV. It operates at high frequencies and constant potential which enables very high quality radiation to be emitted, whatever the conditions of use. The choice of voltage (60 or 70kV) and intensity (4 or 8mA) enables optimal diagnosis of intraoral X-rays with a 35% reduction in exposure time, compared with monophased units. The timer can control up to two DC generators. The extremely short exposure times (minimum 0.02 sec.) are suitable for digital radiology with CCD, CMOS or phosphor plate sensors (radiation is 50% less that that used with “E” type films).
Classification: Electromedical equipment, Class 1 type B
Supply voltage: 230V~ - monophase 50/60Hz
Power absorption at 230V: 1.4kVA
Resistance of the line: 0.5 ohms?
X-ray tube: New Toshiba DG 073B DC tube
High voltage: 60-70kV
Anode current: 4-8mA
Focal spot: 0.7mm
Total filtration: Equivalent to 2mm Al at 70kV
Tube inherent filtration: Equivalent to 0.8mm Al at 70kV
Leakage radiation: Less than 0.25mGy/hour maximum at a distance of 1m from focal spot
Long cone: Focus to cylinder tip distance = 31cm (12”)
Timer: Exposure time can be set from 0.02 to 3.2 seconds
X-ray emission control: “Dead man” switch with 3m spiral cable
Total weight: 25kg (weight of the head: 5.5kg)
Manufactured in compliance with currently applicable regulations and standards (EC Directive 93/42/EEC)
Blue X Intraoral:
• Modern design
• Simple exposure settings
• 3 lengths of arm – from 137–185cm
• Powerful 70kVp radiation beam for superb penetration power
• Range of exposure times from 60ms to 3.2 seconds
• mobile version available
• Small focal spot
OPG Veraview:
• Speed and safety
• 8.2 seconds exposure time – almost half that of other machines
• Range of models available
• Auto focus & exposure
• Conventional and digital versions available also tomographic and cephalometric.
• Ultra low radiation dose – almost half previous models.
• Dental diagnosis has never been faster,safer or better
• Upgradeable